EasyManua.ls Logo

INNO VIEW 5 - Fiber Zoom Function on Screen; Preparing the Fibers

INNO VIEW 5
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Basic Operation 21
Fiber zoom function on screen
Users can zoom in the ber image by double tapping on the screen so as to
observe the splicing result and check the splicing condition.
Preparing the bers
3 Steps should be carried out before splicing:
Step 1: Strip the ber
Remove at least 50mm of secondary coating (valid for both tight and loose tube
secondary coating) and approximately 30~40mm of primary coating with an
appropriate stripper.
Step 2: Clean the ber
Clean bare bers with pure alcohol-soaked gauze or lint-free tissue.
Step 3: Cleave the ber
In order to ensure the best splicing result, cleave the bers with high precision
cleaver such as INNO Instrument VF series ber cleaver, and strictly control the
cleaving lengths shown as below.
Note: Remember to slip a heat-shrink sleeve onto either end of the bers at the
beginning of each ber preparation.
Examples of cleaving lengths
10 10
Use blue V-grooves (clamping on bare ber)
Primary coatingTight secondary coating

Table of Contents