46
Appendix II
List of error messages
During the splice operating process, if the error messages are shown on the
screen, please follow the solution precisely as shown in the list below. If it is not
possible to solve the problem, the splicer may require service by a qualied service
center. In this case, please contact sales agents.
Error Message Reason Solution
Left Fiber Place Error
The ber endface is placed
beyond the electrode centerline.
Press RESET button, and reload
the bers to make sure that
ber endface is placed between
V-groove and the centerline of
electrodes.
Right Fiber Place Error
Press Motor Distance
Over Limit
The ber is not set correctly at the
bottom of the V-groove, which
results in that the ber offsets
beyond motor formation range.
Press RESET button, and
reposition the ber at the bottom
of the V-groove.
Press Motor Error The motor might be damaged. Consult your nearest sales agent.
Search Fiber Endface
Failed
The ber is not set correctly at
the bottom of the V-groove.
Press RESET button, and
reposition the ber at the bottom
of the V-groove.
Arc Failure Arc discharge does not occur.
Conrm the electrodes in
the proper position; replace
electrodes.
Align Motor Distance
Over Limit
The ber is not set correctly at
the bottom of the V-groove.
Press RESET button, and
reposition the ber at the bottom
of the V-groove.
Search Fiber Clad Failed
The ber is not set correctly at
the bottom of the V-groove.
Press RESET button, and
reposition the ber at the bottom
of the V-groove.
Fiber Border Position
Wrong
Dust on the ber surface.
Re-prepare the ber; clean the
lens; redo [Dust Check].