Appendix III 50
AUTO mode is only applicable for SM, MM, NZ bers. AUTO mode may not be
able to identify while splicing special bers.
Mismatch between estimated splice loss and actual splice loss.
The estimated loss is just an estimated number by calculation, for reference
only.
The optical components of the splicer need to be cleaned.
Fiber protection sleeve does not shrink completely.
Extend the heating time.
Method to cancel heating process.
Press HEAT button to stop during heating process. The LED light will go off
after pressing the key.
Fiber protection sleeve sticks to heating plate after shrinking.
Use a cotton swab or a similar soft tip object to push and remove the sleeve.
Forgot password.
Please contact the sales agent.
No arc power change after [Arc Calibration].
It is the internal parameters that are calibrated and adjusted after arc
calibration. The displayed arc power in each splice mode does not change.
Forgot to load bers while execute some specic maintenance function that
bers are needed.
Return key is invalid in this case. Open the windproof cover, load prepared
bers in the splicer, and press [SET] to continue or press [RESET].
Upgrading failure.
When users use the new USB to upgrade, the splice may not be able to identify
the upgrade le correctly. In this case, please re-plug the USB and restart the
splicer.
Check if the upgrade le name and the format are correct.
Please contact sales agent if you still cannot solve the problem.
Other
Please refer to the video in the user CD.