Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:PotentialObstructionAlert
GREENlightilluminated.
Rapid
Beep...Beep...
Beep...Beep
1.Possibleinternalobstruction
intheoxygenpath.Kinked
orblockedtubing,cannulaor
humidier.
2.Flowmetersetat0.5L/minor
less.
1.Inspectforkinksorblockages.Correct,cleanor
replaceitem.Oncecorrected,turnpowerOfffor60
secondsandthenturnpowerbackOn.
2.Resetowmetertoprescribedowrate.Ifprescribed
owrateisbelow0.5L/min,usethePediatric
FlowmeterIRCPF16AW.
TheuseofthePediatricFlowmeterwilldeactivate
thePotentialObstructionAlert.
Alarm:SystemFailure:
YELLOWlightIlluminated.
Shortbeeps,longpause
Beep...........Beep.............
1.Lowoxygenpurity.
2.Kinkedorblockedtubing,
cannulaorhumidier .
3.Flowmetersetat0.5L/minor
less.
1.CleanorReplacelters.
2.Inspectforkinksorblockages.Correct,cleanor
replaceitem.Oncecorrected,turnpowerOfffor60
secondsandthenturnpowerbackOn.
3.Checkowmeterissetto1.0L/minormore.Refer
to10.1Specications,page36
.
GREENlightwithYELLOW
LightFlashing.
1.Internalrepairsrequired.1.Callservicesupplier.
1195056-B35

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V and is the answer not in the manual?

Invacare Perfecto2 V Specifications

General IconGeneral
Oxygen Concentration87% to 95.6%
Flow Rate0.5 to 5 LPM
Power Source120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12.7°C to 32.2°C)
Operating Humidity15% to 95% non-condensing