Invacare®Perfecto₂™VOxygenConcentrator
10TechnicalData
10.1Specications
ElectricalRequirements:
230VAC+10,-15%(253VAC/195.5VAC),50Hz
RatedCurrentInput:
1.6A
SoundPressureLevel:
<
40dBAweightedfor3L/minow
<40dBAweightedfor5L/minow
SoundPowerLevel:
<40dBAweightedfor5L/minow
Altitude:
Upto1828m(6,000ft)abovesealevelwithoutdegradationofconcentrationlevels.
Atmosphericpressurerange:101.33kPa—81.22kPa
OxygenOutputConcentration
Levels:
87%to95.6%at0.5to5L/min
93%±3%at2L/min
Concentrationlevelsachievedafterinitialwarm-upperiod(approximately30
minutes)
MaximumOutletPressure:
34.5kPa±3.45kPa(5psi±0.5psi)
FlowRange:
0.5to5L/min(maximum).Forowrateslessthan1L/min,werecommendtheuseof
theInvacarePediatricFlowmeterAccessory(IRCPF16AW).
PotentialObstructionAlert:
0L/minto0.5L/min
Theconcentratordetectsaconditionthatmayindicateapotentialobstructionofthe
outputoxygen.Rapidaudiblebeepingalert(thisalertisdeactivatedwhenaccessoriesare
connected).Maybeassociatedwithowsettingof0.5L/minorless.
PowerConsumption:
340Wtypicalat230VAC,50Hz
361195056-B