General
DANGER!
RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygendueto
powerfailureordevicemalfunction.
ALWAYShaveabackupsourceofoxygenreadily
available.
Intheeventtheconcentratorfailstoproduce
oxygen,theconcentratorwillbrieyalarm
signalingthepatienttoswitchtotheirbackup
sourceofoxygen.RefertoTroubleshootingfor
moredetail.
1.3IndicationsForUse
Toprovidesupplementaloxygentopatientswithrespiratory
disorders,byseparatingnitrogenfromroomair,bywayofa
molecularsieve.
1.4Description
TheInvacarePerfecto
2
VOxygenConcentratorisusedby
patientswithrespiratorydisorderswhorequiresupplemental
oxygen.Thedeviceisnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasranges
from87%to95.6%.Theoxygenisdeliveredtothepatient
throughtheuseofanasalcannula.
TheInvacarePerfecto
2
VOxygenConcentratorusesa
molecularsieveandpressureswingadsorptionmethodology
toproducetheoxygengasoutput.Ambientairentersthe
device,islteredandthencompressed.Thiscompressed
airisthendirectedtowardoneoftwonitrogenadsorbing
sievebeds.
Concentratedoxygenexitstheoppositeendoftheactive
sievebedandisdirectedintoanoxygenreservoirwhereit
isdeliveredtothepatient.
TheInvacarePerfecto
2
VOxygenConcentratoriscapableof
operationbythepatientinahomeenvironmentorinan
institutionalenvironment.Deviceoperatesatanominal230
VAC/50Hertzsupply.
Serviceinformationwillbeavailableuponrequestto
qualiedtechnicalpersonnelonly.
1.5Contraindications
Therearenoknowncontraindications.
1195056-B
7