Usage
arapidbeepingoftheaudiblealarm.Checkyour
tubingoraccessoriesforblockedorkinkedtubingora
defectivehumidierbottle.Afterratedowisrestored
tobetween0.5L/minand0.75L/min,thePotential
ObstructionAlertwillturnoff.
APotentialObstructionAlertindicatesa
conditionthatmaybeassociatedwithapartial
orcompleteobstructionofoxygenoutput.
Theuseofsomeaccessoriessuchasthe
PreciseR
X
™pediatricowmeterandtheHomeFill
compressorwilldeactivatethePotential
ObstructionAlert.
6.3.5SensO
2
OxygenPurityIndicator
Thisfeaturemonitorsthepurityofoxygengenerated
bytheoxygenconcentrator.Ifpurityfallsbelowfactory
presetstandards,indicatorlightsonthecontrolpanelwill
illuminate.
6.3.6InitialStartupoftheConcentrator
Concentratormaybeusedduringtheinitialwarm-up
time(approximately30minutes)whilewaitingfor
theO
2
puritytoreachmaximum.
Whentheunitisturnedon,theyellowlightwillcomeon
indicatingtheunitisoperatingandisintheinitialwarm-up
period.Whenthegreenlightcomeson,theunitisin
normaloperatingmode(SYSTEMOK/O
2
greaterthan85%).
Onceinnormaloperatingmode,thelightswillbecontrolled
dependingontheoxygenconcentrationvalues.
6.3.7ExplanationofOxygenPurityIndicatorLights
1195056-B
27