Invacare Perfecto2 V - 10.2 Electromagnetic Compliance (EMC)

Invacare Perfecto2 V
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TechnicalData
10.2ElectromagneticCompliance(EMC)
Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11
GroupI
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissions
areverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic
equipment.
RFemissions
CISPR11
ClassB
Harmonic
emissions
IEC61000-3-2
ClassA
Voltageuctuations
/ickeremissions
IEC61000-3-3
Complies
TheDeviceissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevelElectromagneticenvironment–guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±2,4,8kVcontact
±2,4,8,15kVair
±2,4,8kVcontact
±2,4,8,15kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.