TechnicalData
10.2ElectromagneticCompliance(EMC)
Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11
GroupI
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissions
areverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic
equipment.
RFemissions
CISPR11
ClassB
Harmonic
emissions
IEC61000-3-2
ClassA
Voltageuctuations
/ickeremissions
IEC61000-3-3
Complies
TheDeviceissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevelElectromagneticenvironmentguidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±2,4,8kVcontact
±2,4,8,15kVair
±2,4,8kVcontact
±2,4,8,15kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
1195056-B39

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V and is the answer not in the manual?

Invacare Perfecto2 V Specifications

General IconGeneral
Oxygen Concentration87% to 95.6%
Flow Rate0.5 to 5 LPM
Power Source120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12.7°C to 32.2°C)
Operating Humidity15% to 95% non-condensing