EasyManuals Logo

Invacare Perfecto2 V User Manual

Invacare Perfecto2 V
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
InvacarPerfecto₂™VOxygenConcentrator
StandardsandRegulatoryListing:Certiedcomplyingwith:
IEC/EN60601-1
IEC/EN60601-1-2
IEC/EN60601-1-6
IEC/EN60601-1-11
ISO80601-2-69
CEmarkedmodels:
IRC5PO2VAW
Electrical:Noextensioncords.
Placement:
Nocloserthan30.5cm(12in)fromanywall,furniture,draperies,orsimilarsurfacesto
assuresufcientairow.
Avoiddeeppilecarpetsandheaters,radiatorsorhotairregisters.
Floorlocationonly.
Noconnedspaces(Example:Noclosets).
Environment:
Smokeandsoot-free
Tubing:
2.1m(7ft)cannulawithamaximum15m(50ft)ofCrush-ProofTubing
(DONOTpinch)
TimeofOperation:Upto24hoursperday
RecommendedStorageand
ShippingTemperature:
-29°Cto65°C(-20°Fto150°F)atupto95%relativehumidity
MinimumOperatingTime:30Minutes
Measurementuncertaintyisincludedinthedevicespecication.AllconditionsatSTPD(standardtemperatureand
pressuredry).
381195056-B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V and is the answer not in the manual?

Invacare Perfecto2 V Specifications

General IconGeneral
Oxygen Concentration87% to 95.6%
Flow Rate0.5 to 5 LPM
Power Source120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12.7°C to 32.2°C)
Operating Humidity15% to 95% non-condensing