Maintenance
3.Thoroughlydrythelterandinspectforfraying,
crumbling,tearsandholes.Replacelterifanydamage
isfound.
4.Reinstallthecabinetlter.
7.4HumidierCleaningandThermicDisinfection
Cleananddisinfecttheoxygenhumidierdailyto
reducelimestonedepositsandeliminatepossible
bacterialcontamination.Followtheinstructions
providedbythemanufacturer .Ifnoneareprovided,
followthesesteps:
1.Washhumidierinsoapywaterandrinsewitha
solutionoftenpartswaterandonepartvinegar .
2.Rinsethoroughlywithhotwater.
3.Airdrythoroughly.
Tolimitbacterialgrowth,airdrythehumidier
thoroughlyaftercleaningwhennotinuse.Refer
to6.3SettingUptheConcentrator,page20foruse.
7.5CleaningandDisinfectionBetweenPatients
WARNING!
RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto
concentrator:
Onlyqualiedpersonnelshouldperform
cleaninganddisinfectionoftheoxygen
concentratorandaccessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossible
pathogenexchangebetweenpatientsdueto
contaminationofcomponentsoraccessories.
Preventivemaintenanceshouldalsobeperformed
atthistimeifnecessary.
1.Disposeofandreplaceallpatientsideaccessoriesnot
suitableformultiplepatientuse,includingbutnot
limitedto:
NasalCannulaandTubing
Mask
Humidier
2.Performmaintenanceproceduresdescribedinthis
manualanditemsonPreventiveMaintenanceChecklist.
3.Checkconcentratorforpossibleexternaldamageor
signsthatitmayrequireserviceorrepair .
4.Ensureconcentratorfunctionsproperlyandallalarms
areinworkingorder .
5.Beforerepackaginganddistributiontonewpatient,
ensurepackagingcontentscontaintheconcentrator,
usermanual,powercord,airinletscoop,assembly
instructions,humidier,cannulaandlabels.
1195056-B31

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V and is the answer not in the manual?

Invacare Perfecto2 V Specifications

General IconGeneral
Oxygen Concentration87% to 95.6%
Flow Rate0.5 to 5 LPM
Power Source120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12.7°C to 32.2°C)
Operating Humidity15% to 95% non-condensing