Invacare®Perfecto₂™VOxygenConcentrator
9Troubleshooting
9.1Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:
MainPowerLoss:
Nolightsilluminated.
ShortBeeps,longpause
Concentratornotoperating,
powerswitchOn.
Beep...Beep...
1.Powercordnotpluggedin.
2.Nopoweratoutlet.
3.Trippedcircuitbreaker .
1.Insertplugintooutlet.
2.Inspecthousecircuitbreakersorfuses.Ifproblem
recurs,useadifferentoutlet.
3.Push/resetcircuitbreaker .Ifproblemrecurs,call
servicesupplier.
Alarm:SystemFailure:
REDlightilluminated.
Continuous
Concentratornotoperating,
powerswitchOn.
Beep...
1.Unitoverheatingdueto
blockedairintake.
2.Insufcientpoweratoutlet.
3.Internalrepairsrequired.
1.Dooneorbothofthefollowing:
a.Removeandcleancabinetlters.
b.Moveoxygenconcentratoratleast30.5cm(12
in)awayfromwalls,draperiesorfurniture.
2.DONOTuseextensioncords.Movetoanother
electricaloutletorcircuit.
3.Callservicesupplier.
341195056-B