TechnicalData
PortableandmobileRFcommunicationsequipment
shouldbeusednoclosertoanypartofthe
Deviceincludingcables,thantherecommended
separationdistancecalculatedfromtheequation
applicabletothefrequencyofthetransmitter.
Recommendedseparationdistance:
ConductedRF
IEC61000-4-6
3Vrms,6Vrms3Vrms,6Vrms
d=1.2√P150kHzto80MHz
RadiatedRF
IEC61000-4-3
10V/m10V/m
d=0.35√P
d=0.7√P
80MHzto800MHz
800MHzto2,5GHz
wherePisthemaximumoutputpowerrating
ofthetransmitterinwatts(W)accordingto
thetransmittermanufactureranddisthe
recommendedseparationdistanceinmetres(m).
FieldstrengthsfromxedRFtransmitters,as
determinedbyanelectromagneticsitesurvey,
a
shouldbelessthanthecompliancelevelineach
frequencyrange.
b
Interferencemayoccurinthevicinityofequipment
markedwiththefollowingsymbol:
a
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,
amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.Toassess
theelectromagneticenvironmentduetoxedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthe
1195056-B
41

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V and is the answer not in the manual?

Invacare Perfecto2 V Specifications

General IconGeneral
Oxygen Concentration87% to 95.6%
Flow Rate0.5 to 5 LPM
Power Source120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12.7°C to 32.2°C)
Operating Humidity15% to 95% non-condensing