CALibration Subsystem
Copyright © Itech Electronic Co., Ltd. 109
Command Syntax
CALibrate:VOLTage:POINt <point>
Parameters
P1 | P2 | P3 | P4
Examples
CAL:VOLT:POIN P2
Related Commands
CAL:STAT CAL:SAV
CALibrate:VOLTage[:LEVel]
This command is only used in calibration mode. It enters a calibration voltage
value that you obtain by reading an external meter. You must first select a
calibration level (with CALibrate:VOLTage:POINt) for the value being entered.
These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE.
Command Syntax
CALibrate:VOLTage[:LEVel] <NRf>
Parameters
<external reading>
Unit
V (volts)
Examples
CAL:VOLT 3.2223
Related Commands
CAL:STAT CAL:SAV
CALibrate:VOLTage:METEr:POINt
This command can only be used in calibration mode. It is used to set the
calibration points of constanrd voltage mode. P1, P2 is used in low voltage
range, P3, P4 is used in high voltage range. It can only use calibrate voltage
meter.

Table of Contents

Other manuals for ITech IT8700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8700 Series and is the answer not in the manual?

ITech IT8700 Series Specifications

General IconGeneral
BrandITech
ModelIT8700 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals