SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 82
Returned Parameters
<NR3>
Related Commands
VOLT:RANG
[SOURce:]VOLTage:RANGe
This command sets the voltage range of the electronic load module. There are
only one voltage ranges.
Command Syntax
[SOURce:]VOLTage:RANGe <NRf+>
Parameters
MINimum through MAX | MINimum | MAXimum | DEFault
Unit
V (volts)
*RST Value
MAXimum
Examples
VOLT:RANG 15
Query Syntax
[SOURce:]VOLTage:RANGe?
Returned Parameters
<NR3>
Related Commands
VOLT
[SOURce:]VOLTage:RANGe:AUTO[:STATe]
This command sets the voltage auto range state of the electronic load module.
Command Syntax
[SOURce:]VOLTage:RANGe:AUTO[:STATe] <bool>