EasyManua.ls Logo

ITech IT8700 Series - [Source:]Current:slew:negative; [Source:]Current:protection:state

ITech IT8700 Series
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 75
[SOURce:]CURRent:SLEW:NEGative
This command sets the slew rate of the current for negative going transitions.
MAXimum sets the slew to the fastest possible rate. MINimum sets the slew to
the slowest rate.
Command Syntax
[SOURce:]CURRent:SLEW:NEGative <NRf+>
Parameters
MINimum to MAXimum | MAXimum | MINimum | DEFault
Unit
A (amps per micro second)
*RST Value
MAXimum
Examples
CURR:SLEW:NEG MAX
Query Syntax
[SOURce:]CURRent:SLEW:NEGative?
Returned Parameters
<NR3>
Related Commands
CURR:SLEW
[SOURce:]CURRent:PROTection:STATe
This command enables or disables the over-current protection feature.
Command Syntax
[SOURce:]CURRent:PROTection:STATe <Bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF