MONITOR DESCRIPTION
Model 3000T Operation Manual
15
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may
lead to a reduce level of safety of the resulting system. Consideration relating to the choice shall include:
Use of the accessory in the PATIENT VICINITY
Evidence that the safety certification of the ACCESSORY has been performed in accordance to the
appropriate IEC 60601-1 and / or IEC 60601-1-1 harmonized national standard.
Model 3000T
Rear Panel:
Fuse Ratings
The fuses are located behind the cover of the power entry module. To replace the fuses, unplug the ac power cord.
Remove the power entry module cover and replace the fuse(s) only with same type and rating T.5A, 250V (Metric
5x20mm).
ECG X1000
and
Synchronized
Out
u
BNC connector
Synchronized
Output
AUXILIARY
(Optional)
Line Voltage
Selector
Power entry
module
PEQ
Ground
Fuses
(Behind the
anel
Serial Number
Label