MAINTENANCE AND CLEANING
Model 3000T Operation Manual 33
MAINTENANCE AND CLEANING
The Monitor
When necessary, clean the exterior surfaces of the monitor with a cloth or swab dampened with a warm water and
mild detergent solution. Do not allow liquids to enter the interior of the instrument.
CAUTION:
Do not autoclave, pressure sterilize, or gas sterilize the monitor.
Do not soak or immerse in any liquid.
Use cleaning solution sparingly. Excessive solution can flow into the monitor and cause damage to internal
components.
Do not touch, press or rub the display and covers with abrasive cleaning compounds, instruments, brushes,
rough surface materials, or bring them into contact with anything that could scratch the display or the
covers.
Do not use petroleum based or acetones solutions, or other harsh solvents, to clean the monitor.
Patient Cables
Do not autoclave the patient cables.
Wipe the cables using a mild detergent solution. Never submerge the cables in any liquid or allow liquids to enter
the electrical connections.
Preventive Maintenance
ECG
Check before connecting the monitor to a new patient that:
Cables and Leads are clean and intact.
The LEAD OFF message is displayed when the patient cable is connected, but the patient leads are not
connected. Connecting the patient leads together should make the message disappear.
The BNC Interconnect cable is clean and intact.
NOTE: There are no user serviceable items contained in the Model 3000T.