EasyManuals Logo

Ivy Biomedical Systems 3000T User Manual

Default Icon
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
MONITOR SETUP
Model 3000T Operation Manual 17
To set the Trace Speed
1. Press the [SETUP] key in the main menu.
2. Press the [ECG] key.
3. Press the [OPTIONS] key.
4. Press the [SPEED] key to select the trace speed. Selections are 25, and 50 mm/s.
NOTE: The [SPEED] key also changes the speed of the recorder.
Default Settings
To reset the monitor to the default settings, turn monitor off by pressing the STBY key; then press and hold the
fourth and fifth soft key (from left to right) while applying power to the monitor by pushing the ON key.
Setting Initial Default
Initial Language Setting English
ECG Size 10mm
Lead II
Trigger output/mark ON
ECG Notch filter ON
Impedance (3000T-B) Enabled
Impedance Threshold (3000T-B) 50k
Impedance Auto (3000T-B) OFF
Pacer Detect Disabled
P-Lock Enabled
Heart Rate Low Limit 60
Heart Rate High Limit 120
Trace Speed 25mm/sec
Recorder Direct
QRS Volume OFF
Alarm Volume High
Alarms OFF
Some settings (see list below) are stored in non-volatile memory which means that the monitor powers up with the
same options that were in effect when power was last turned off.
Setting Options
Speed 25mm/sec 50mm/sec
Recorder Direct Timed Delay
Alarm Vol. High Low
P-Lock Enabled Disabled

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 3000T and is the answer not in the manual?

Ivy Biomedical Systems 3000T Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model3000T
CategoryMedical Equipment
LanguageEnglish

Related product manuals