JP-33 SERVICE MANUAL
- -
10
Maintenance Program - Change Chemical
When & How to Replace Chemicals
Important: The chemistry’s condition is important for proper film processing. Exhausted and/or
contaminated chemistry will lead to poor images. Since the processor is used in the final stage of
generating an image, it is normally blamed for poor image quality, overlooking the chemistry.
Before assuming the poor image quality is due to a processor problem we highly recommend that
the chemistry be checked.
1. Generals
When is to replace the Chemistry.
: Refer to trouble shooting guide section 4, reasons for poor image quality.
When there are no problems, chemistry should be replaced in accordance to the chemical
manufacturer’s guidelines.
2. Procedure for replacing the chemistry
1) Removing old chemistry.
- Before opening drain valves, make sure the drain hoses are inserted inside the tank or drain.
- Remove all the racks (DEV, FIX, WASH). Be careful worm wheel (gear) is lubricated.
DO NOT GET LUBRICANT ON CLOTHING OR OTHER PARTS OF THE RACK.
Open top cover Æ remove rack cover Æ open rack locks Æ remove DEV, FIX, WASH
racks Æ be careful not to spill chemicals on the floor, on any electrical components and
keep the chemicals away from the operators body and clothes.
- Open drain valve to drain the old chemicals.
2) Waste disposal
Disposal of waste should be performed in accordance to local rules and regulations.