10 Appendix
10.1 Replacing the battery
When the battery charge indicator shows that the battery is nearly empty
(), you should replace it.
Type of battery LR6 - alkaline battery (to IEC and JIS)
Number of
batteries
one battery
Battery life With a battery, the TDA-300(0) can measure temperatures continuously for
about 400 hours (ambient temperature 23 °C). The effective battery life
depends on the operating and ambient conditions.
Displayed data
Configuration
data
A
When the battery is replaced, the instrument is started in the
RESET condition. For this reason, there are some parameters
which will be initialized (factory setting). In addition, please note
that all recorded data will be deleted on the TDA-300.
Parameter TDA-300 TDA-3000
Peak value is reset to the first measurement that is recorded after
the reset
Bottom value is reset to the first measurement that is recorded after
the reset
Temperature data are deleted are retained
Parameter TDA-300 TDA-3000
TAG filter is reset (switched off)
TAG name is retained
Limits (high/low) are retained
Interval is retained
Selected user No. --- is set to 1
User names --- are retained
Language is retained
Date format is retained
Sensor type is retained
Temperature unit is retained
Decimal place is retained
Temperature correction is retained