KANE905 Operators Manual Page 12
To obtain the quoted specification an instrument should be calibrated
with clean ambient air at standard temperature and pressure (STP).
Once the time has reached zero
an audible beep will be heard and
will show the selected fuel on the
following display:-
Press
This zeros the toxic sensor and
sets Oxygen to 20.9%. The next
screen is the MAIN DISPLAY of
the analyser:
Use and to change the display.
All parameters are detailed in APPENDIX A - MAIN DISPLAY
PARAMETERS.
MAIN DISPLAYS
The main display can be changed to show either 4 or 8 parameters at
one time.
Two options are available when 4 parameters are selected.
4 Page Mode displays 4 lines of data in set format, each page is
predefined.
Line scroll mode allows you to customise the display to show the
data you require.
8 Page Mode displays 8 parameters on 4 lines in set format, the
bottom two can be changed.
Changing between the different modes is detailed in DISPLAY MENU .