KANE905 Operators Manual Page 41
APPENDICES
A. MAIN DISPLAY PARAMETERS
The parameters and their meanings are detailed as follows : -
DATE: Analyser date. See SET UP MENU to change
TIME: Analyser time. Use SET UP MENU to change
BATTERY:
(BAT)
Displays the battery level from 0-100%. The analyser will
flash RECHARGE BATTERY at less than 10% of charge.
With the charger connected the display shows AC ON.
NETT:
(T)
Nett temperature calculated by deducting the internal
AMBIENT temperature from the measured FLUE
temperature. Displays in either C (C) or F (F) and will
display NOT FITTED (N/F) if flue probe is not connected.
If an external INLET probe is used then INLET is deducted
from FLUE.
O2:
Oxygen reading in percentage %.
CO:
Carbon Monoxide reading indicated in ppm or mg/m3. If
the figures are referenced to oxygen then the display will
show CO(n). See SELECT menu 5.2.2 for oxygen
reference. The display will read ‘O2 > 20%’ if referenced
values selected and instrument is in clean ambient air.
EFF (G):
Combustion Efficiency calculation displayed in
percentage. Gross G or Net N can be set see SELECT
menu 5.2.3. The calculation is determined by fuel type see
Appendix B for calculation. The efficiency is displayed
during a combustion test, ‘- -’ is displayed while in fresh
air.
CO2:
Carbon Dioxide calculation determined by the type of fuel.
This only shows a reading when a combustion test is
being carried out. ‘- -’ is displayed while in fresh air.
FLUE:
(Tf)
Temperature measured by flue gas probe in Centigrade or
Fahrenheit. Will show ambient temperature after fresh air
calibration and NOT FITTED (N/F) or FAULT (FLT) if
probe disconnected.