KANE905 Operators Manual Page 22
level in the flue and hence dilute any toxic gas reading.
Oxygen referencing gives readings as if they were undiluted.
SMOKE: Allows the user to enter a smoke test number from 0-9. This
value will be printed on the standard printout. Default value is
OFF.
RESET : Allows the user set the Oxygen to 20.9% and zero the toxic
sensors without turning the analyser off.
Selecting YES and will
display the following screen.
After pressing the analyser will count down for 10
seconds and then return to the main display.
WARNING: The sensors must only be reset if you are sure
they have been sampling fresh air for at least 3 minutes.
Errors in measurement will occur if the sensors are reset
during or just after sampling.
PITOT: When pressure units set to m/s Pitot Mode is active, adjust
the PITOT setting/value here.
SUB MENU - 2. UNITS
Allows all displayed units to be changed.
TEMP: Choose from Centigrade, C, or Fahrenheit, F.
GAS: Changes the toxic gas measurement units. Select from
volumetric readings, parts per million (ppm) or mass flow
reading milligrams per cubic meter (mg/m
3
). When set to
m/s. Pitot flow mode is active. FLOW will show on measure
screen instead of XAIR.