Instructions for use ELECTROmatic M/C and PM/PC
6 Operation | 6.1 Switching the ELECTROmatic on/off
54 / 78
6 Operation
CAUTION
Germ formation.
Infections.
Before treating a patient, let the spray air and spray water exit for at least
20 seconds.
Before the first start-up and after downtimes (weekends, holidays, vaca-
tions, etc.), the air and water lines must be purged and/or rinsed.
The germ reduction of the ELECTROmatic must be carried out via the
dental unit.
See also:
2 8 Reprocessing steps in accordance with DIN EN ISO 17664, Page 61
NOTICE
Incorrectly set parameters.
Property damage from incorrect input values.
Check all input values on the display before use.
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
Note
Please note the transmission/reduction ratio of the attachment handpieces as
these have an impact on the displayed speed of the clamped tool.
6.1 Switching the ELECTROmatic on/off
The unit is ready for operation as soon as it is connected.
To switch the product on, connect the device to the power circuit.
ð An LED on the rear of the ELECTROmatic M/C flashes.
ð The treatment parameters are displayed on the display of the
ELECTROmatic PM/PC:
To switch the device off, unplug the unit from the mains.
ð The LED on the rear of the ELECTROmatic M/C is off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KaVo ELECTROmatic M and is the answer not in the manual?

KaVo ELECTROmatic M Specifications

General IconGeneral
BrandKaVo
ModelELECTROmatic M
CategoryDental equipment
LanguageEnglish

Related product manuals