39
:TEMPerature
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
:TCouple
:TYPE <name>
:TYPE?
:RJUNction
:RSELect <name>
:RSELect?
:SIMulated <n>
:SIMulated?
:REAL
:TCOefficient <n>
:TCOefficient?
:OFFSET <n>
:OFFSet?
Select filter type: (MOVing or
REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure thermocouple:
Select thermocouple type (J, K,
or T).
Query thermocouple type.
Path to configure reference
junction:
Select reference type (SIMu-
lated or REAL).
Query reference type.
Specify simulated tempera-
ture in °C (0 to 50).
Query simulated tempera-
ture.
Path to configure real refer-
ence junction:
Specify temp coefficient
(-0.09999 to 0.09999).
Query temp coefficient.
Specify voltage offset at
0°C (-0.09999 to
0.09999).
Query voltage offset.
Table 10. SENSe command summary (cont.)
Command Description
2000-903-01A Page 39 Thursday, December 23, 1999 7:52 AM