Performance Verification
1-12
1.8.4 AC current verication
AC current verification is performed by applying accurate
AC currents at specific frequencies and then verifying that
Model 2002 readings fall within specified limits.
Follow the steps below to verify ACI measurement accuracy.
CAUTION
Do not apply more than 2A, 250V to the
AMPS input, or the current protection
fuse will blow.
1. Connect the Model 2002 to the calibrator, as shown in
Figure 1-5. Be sure to connect calibrator HI to the
AMPS input, and connect calibrator LO to INPUT LO
as shown.
2. Turn on the Model 2002 and the calibrator, and allow a
one-hour warm-up period before making measure-
ments. Be sure the calibrator is set for normal current
output.
3. Restore Model 2002 factory default conditions, as
explained in paragraph 1.7.
4. Select the AC current function and the 200µA range on
the Model 2002.
5. Set the calibrator output to 190.000µA AC at a fre-
quency of 40Hz, and allow the reading to settle.
6. Verify that the Model 2002 reading is within the limits
for the present current and frequency summarized in
Table 1-7.
7. Repeat steps 5 and 6 for each frequency listed in Table
1-7.
8. Repeat steps 5 through 7 for the remaining ranges and
frequencies listed in Table 1-7.
Figure 1-5
Connections for AC current verification
2002 MULTIMETER
Output HI
Input
LO
Output
LO
Model 2002
5700A Calibrator (Output AC Current)
Amps
Note: Be sure calibrator is set for
normal current output.
190.000 µAAC RMS

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A