EasyManua.ls Logo

Keithley 2002

Keithley 2002
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
2-13
AC CALIBRATION PHASE
Disconnect inputs; press ENTER
6. Press ENTER to begin AC calibration, which will take
about six minutes to complete. During AC calibration,
the instrument will display the following:
Calibrating AC: Please wait
7. Once the process has been successfully completed, the
message below will be displayed, and you can press
ENTER or EXIT to return to normal display:
AC CAL COMPLETE
Press ENTER or EXIT to continue.
2.9.2 IEEE-488 bus AC self-calibration
Procedure:
1. Disconnect all test leads and cables from the INPUT and
SENSE jacks.
2. Send the following command over the bus: ":CAL:UN-
PR:ACC".
3. Wait until calibration has been completed before send-
ing any further commands (about six minutes).
4. Check for calibration errors by using the :SYST:ERR?
query.
2.10 Low-level calibration
Low-level calibration is normally performed only at the fac-
tory when the instrument is manufactured and is not usually
required in the field. The following paragraphs give detailed
procedures for performing low-level calibration should it
ever become necessary in the field.
NOTE
Low-level calibration is required in the
field only if the Model 2002 has been
repaired, or if the other calibration proce-
dures cannot bring the instrument within
stated specifications. The low-level cali-
bration procedure includes the compre-
hensive calibration steps discussed in
paragraph 2.8.
2.10.1 Recommended equipment for low-level
calibration
Table 2-7 summarizes recommended equipment for low-
level calibration. Alternate equipment may be used as long as
corresponding specifications are at least as good as those
listed in the table.
2.10.2 Low-level calibration summary
Table 2-8 summarizes the steps necessary to complete the
low-level calibration procedure. The procedure should per-
formed in the order shown in the table. Calibration com-
mands shown are to be used when calibrating the unit over
the IEEE-488 bus.
WARNING
Some low-level calibration steps require
the use of hazardous voltages.
See Appendix B for example calibration programs.

Table of Contents

Other manuals for Keithley 2002

Related product manuals