EasyManuals Logo

Keithley 2002 Calibration Manual

Keithley 2002
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
Calibration Command Reference
3-5
3.3.6 :DATE (:CALibration:PROTected:DATE)
Purpose To send the calibration date to the instrument.
Format :cal:prot:date <yr>,<mon>,<day>
Parameters <yr> = year (yyyy, 1993 to 2092)
<mon> = month (mm, 1 to 12)
<day> = day of month (dd, 1 to 31)
Query Format :cal:prot:date?
Response <yr> , <mon> , <day>
Description The :DATE command allows you to store the calibration date in instrument memory for future
reference. You can read back the date from the instrument over the bus by using the :DATE?
query, or by using the CALIBRATION selection in the front panel menu.
Programming Note The year, month, and day parameters must be delimited by commas.
Examples :CAL:PROT:DATE 1994,12,16 Send cal date (12/16/94).
:CAL:PROT:DATE? Request date.
3.3.7 :NDUE (:CALibration:PROTected:NDUE)
Purpose To send the next calibration due date to the instrument.
Format :cal:prot:ndue <yr>, <mon>, <day>
Parameters <yr> = year (yyyy, 1993 to 2092)
<mon> = month (mm, 1 to 12)
<day> = day of month (dd, 1 to 31)
Query Format :cal:prot:ndue?
Response <yr>, <mon>, <day>
Description The :NDUE command allows you to store the date when calibration is next due in instrument
memory. You can read back the next due date from the instrument over the bus by using the
:NDUE? query, or by using the CALIBRATION-DATES selection in the front panel menu.
Programming Note The next due date parameters must be delimited by commas.
Examples :CAL:PROT:NDUE 1995,12,16 Send due date (12/16/95).
:CAL:PROT:NDUE? Request due date.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A