EasyManua.ls Logo

Keithley 2002 - Page 40

Keithley 2002
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
2-11
(Send the actual calibration value in the range of 9.5V to
20.5V if you are using a different voltage.)
7. Wait until the Model 2002 finishes this step before going
on.
Step 4: Ohms calibration
1. Set the calibrator output to 1MΩ, and turn external sense
on.
NOTE
External sensing (4-wire ohms) should be
used when calibrating all resistance
ranges.
2. Send the following command to the Model 2002:
:CAL:PROT:DC:OHM1M <value>
Here, <value> is the actual calibrator resistance value. For
example, if the calibrator resistance is 1.002MΩ, the com-
mand would appear as follows:
:CAL:PROT:DC:OHM1M 1.002E6
(The allowable range for this parameter is from 475E3 to
1.025E6.)
3. Wait until the Model 2002 finishes 1MΩ calibration be-
fore continuing.
4. Set the calibrator resistance to 100kΩ, and make sure
external sense is still turned on.
5. Repeat steps 2 and 3 for each of the remaining ohms cal-
ibration points as shown in Table 2-5.
Step 5: Amps calibration
1. Connect the calibrator to the AMPS and INPUT LO
jacks, as shown in Figure 2-3.
2. Set the calibrator output to 200.000µA, and place the
unit in operate.
3. Send the following command to the Model 2002:
:CAL:PROT:DC:A200U 200E-6
If you are using a different calibration value, be sure to sub-
stitute that value for the parameter shown above. (The allow-
able range is from 95µA to 205µA.)
Wait for the instrument to complete this step before
continuing.
4. Repeat steps 2 and 3 for the remaining amps calibration
points shown in Table 2-6.
Step 6. Open-circuit calibration
1. Disconnect all test leads from the Model 2002 INPUT
and AMPS jacks.
2. Send the following command to the instrument:
:CAL:PROT:DC:OPEN
3. Wait until open-circuit calibration is complete before
going on to the next step.
Table 2-5
Ohms calibration summary
Cal
point
Nominal
resistance Allowable range Command*
2MΩ
200kΩ
20kΩ
2kΩ
200Ω
20Ω
1MΩ
100kΩ
19kΩ
1.9kΩ
190Ω
19Ω
475kΩ to 1.025MΩ
95kΩ to 205kΩ
9.5kΩ to 20.5kΩ
0.95kΩ to 2.05kΩ
95Ω to 205Ω
9.5Ω to 20.5Ω
:CAL:PROT:DC:OHM1M 1E6
:CAL:PROT:DC:OHM200K 100E3
:CAL:PROT:DC:OHM20K 19E3
:CAL:PROT:DC:OHM2K 1.9E3
:CAL:PROT:DC:OHM200 190
:CAL:PROT:DC:OHM20 19
* Nominal resistance values shown. Use exact calibrator resistance value for command parameter.

Table of Contents