Section
3: Functions and features Series 2600B System SourceMeter® Instrument
Reference Manual
3-60 2600BS-901-01 Rev. B / May 2013
Output characteristics:
In addition to trigger events from other trigger objects, the digio.trigger[N].assert() and
tsplink.trigger[N].assert() commands generate a low pulse that is similar to the falling
edge trigger mode.
Figure 76: RisingA output trigger
Either edge trigger mode
The either edge trigger mode generates a low pulse and detects both rising and falling edges.
Input characteristics:
All rising or falling edges generate an input trigger event.
Figure 77: Either edge input trigger