Series 2600B
System SourceMeterĀ® Instrument Reference Manual Appendix B:
Calibration
2600BS-901-01 Rev. B / May 2013 B-9
3. Capture the reported voltage measurement.
4. Calculate the current from measured voltage and characterized resistance.
5. Verify output current accuracy for each of the currents for the 1 nA to 100 nA ranges listed in the
table titled "Model 2634B/2635B/2636B output current accuracy limits."