EasyManuals Logo
Home>Keithley>Measuring Instruments>2651A

Keithley 2651A User Manual

Keithley 2651A
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Section 5: Capturing high power pulse waveforms Model 2651A High Power System SourceMeter® Instrument User's Manual
Figure 12: GPIB communication example for fast ADC
CURSOR
SRC MEAS LIMIT MODE
POWER
AUTO
O
U
T
P
U
T
O
N
/
O
F
F
2651A HIGH POWER SYSTEM SourceMeter
®
12 3
4
DISPLAY
CONFIG
56
78 9
0
+/-
0000
SPEED
DIGITS
FILTER
REL
RECALL
STORE
MENU
TRIG
ENTER
LOCAL
EXIT
RUN
LOAD
CONFIG
NF
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
P
U
S
H
T
O
E
D
I
T
/
E
N
T
E
R
+3.21000V
ARM
SrcA:+20.0000 A LimAL10.0000V
GPIB
Controller
Device connections
Make the connections as illustrated in the following figure. To ensure good contact, use care when
making connections.
Figure 13: Connections for fast ADC example
GGGG
S
LO
SENSE/GUARD
S
HI
LO HI
OUTPUT
40V, 50A MAX.
S HI
HI
S LO
LED
(DUT)
LED
(DUT)
LO
Model 2651A
Equivalent
HI
S HI
LO
S LO
SMU
Configuring the trigger model
To output precisely timed high power pulses and accurately capture the pulse with the fast ADC, the
trigger model of the Model 2651A High Power System SourceMeter
®
Instrument must be used. Using
the trigger model, we can output pulses as large as 50 A with pulse widths as small as 100 μs and
position our measurements with microsecond resolution. The image below is a diagram of the trigger
model used in this example.
5-2 2651A-900-01 Rev. A / March 2011

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals