Section 7: Increasing SMU current sourcing ability   Model 2651A High Power System SourceMeter® Instrument User's Manual 
The following script contains all the code necessary to perform an Rds(on) sweep up to 100 A using 
two Model 2651A High Power System SourceMeter
®
 Instruments and a Series 2600A System 
SourceMeter
®
 Instrument. This script: 
•  Initializes the TSP-Link
®
 connection 
•  Configures the source-measure units (SMUs) 
•  Configures the trigger models of the two Model 2651A instruments 
•  Prepares the reading buffers 
•  Executes the test 
•  Stores the collected data in a format that can be copied from the instrument console and pasted 
into a Microsoft
®
 Excel
®
 spreadsheet. 
 
The script is written using Test Script Processor (TSP
®
) functions rather than as a single block of 
inline code. TSP functions are similar to functions in other programming languages, such as 
Microsoft
®
 Visual C
®
 or Visual Basic
®
, and must be called before the code contained in them is 
executed. Because of this, running the script alone will not execute the test. To execute the test, run 
the script to load the functions into test script memory and then call the functions. Refer to the 
documentation for Test Script Builder or TSB Embedded for directions on how to run scripts and enter 
commands using the instrument console. 
The script contains comments that describe what is being performed by the lines of code and 
document the functions in the script. Lines starting with node[2]. are commands that are being sent 
to Model 2651A on node 2 through the TSP-Link interface. Lines starting with node[3]. are 
commands that are being sent to the Series 2600A through the TSP-Link interface. All other 
commands are executed on the master Model 2651A on node [1]. 
--[[
   Title:   Combining SMUs for 100A Example
   Description:   This script is designed to perform an Rds(on)sweep on a power
   MOSFET device. It combines two 2651A SMUs in parallel to perform a current
   sweep up to 100A. Data collected from the sweep is then returned in a
   Microsoft Excel compatible format for plotting and analysis.
   Equipment needed:
      2x 2651A
      1x 26xxA
      2x TSP-Link Cable
   TSP-Link Configuration:
   -----------------------
   Unit         |   Node #
   2651A #1     |   1
   2651A #2     |   2
   26xxA        |   3
   Master Node (PC Interface): Node 1
]]
 
 
 
7-10  2651A-900-01 Rev. A / March 2011