4: TSP command reference DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
4-6 DMM6500-905-01 Rev. E June 2021
cal.lock()
This function prevents access to instrument calibration.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
cal.lock()
Details
Calibration data is locked during normal operation. To perform calibration, you must unlock
calibration.
This command does not save calibration data.
Calibration data is lost if it is you do not save it before sending cal.lock(). Use cal.save() to
save the data.
An error is generated if this command is issued when calibration is already locked.
Example
cal.unlock("KI000CAL")
-- Perform operations to generate the calibration data
cal.save()
cal.lock()
Unlock the calibration for the instrument using the default password.
Save the calibration data.
Lock the calibration data.
Also see
cal.save() (on page 4-8)
cal.unlock() (on page 4-9)

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings

Related product manuals