4: TSP command reference DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
4-10 DMM6500-905-01 Rev. E June 2021
cal.verify.date
This attribute contains the date of the last calibration verification.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
Not applicable
Usage
verifyDate = cal.verify.date
cal.verify.date = os.time({year = year, month = month, day = day})
cal.verify.date = os.time({year = year, month = month, day = day, hour = hour, min =
minute, sec = second})
verifyDate
The date when the last verification occurred
year
Year; must be more than 1970
month
Month (1 to 12)
day
Day (1 to 31)
hour
Hour in 24-hour time format (0 to 23)
minute
Minute (0 to 59)
second
Second (0 to 59)
Details
The date and time is returned in the format:
MMM DD YYYY HH:MM:SS.NNN
Where:
MMM DD YYYY is the month, date, and year
HH:MM:SS.NNN is the hour, minute, second, and fractional second
You can read this command if calibration is locked or unlocked. To set the date and time, calibration
must be unlocked.
When using the os.time() function, if no parameters are specified, the current date and time of the
instrument is used.
The verification date is also available from the front panel:
1. Press the MENU key.
2. Select Calibration.
3. The verification date is displayed as the Calibration Date.

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link