-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 2:
Performance verification
DMM6500-905-01 Rev. E June 2021 2-41
Capacitance verification
To compensate for capacitance offset of the cable and 1 µF thru 100 µF decade box:
1. Connect the DMM6500, shielded banana cable, banana to dual BNC cable, and 1 µF through
100 µF decade capacitor box as shown in the following diagram.
Figure 15: Connect DMM6500 to decade capacitor box