Section
3: Adjustment DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
3-2 DMM6500-905-01 Rev. E June 2021
Environmental conditions
To make sure you get accurate results, the environment must meet the following conditions.
Temperature and relative humidity
Conduct the adjustment procedures in a test environment with:
A stable ambient temperature controlled to vary less than ±1 °C during the period of adjustment.
Keithley Instruments recommends a calibration adjustment temperature (T
CAL
) of 23 °C. If a
different nominal temperature is used, it should be noted on the calibration report. A relative
humidity of less than or equal to 40 percent, unless otherwise noted.
No direct airflow on the input terminals.
Line power
The DMM6500 requires a line voltage of 100 V to 240 V and a line frequency of 400 Hz, 50 Hz or 60
Hz.
Calibration adjustments should be performed within this range.
The instrument automatically senses the line frequency at power-up.
Warmup period
Allow the DMM6500 to warm up for at least 30 minutes before conducting the adjustment procedures.
If the instrument has been subjected to temperature extremes (more than 5 °C above or below T
CAL
),
allow additional time for the internal temperature of the instrument to stabilize. Typically, allow an
additional 30 minutes to stabilize an instrument that is 10 °C outside the specified temperature range.
Also allow the test equipment to warm up for the time recommended by the manufacturer.

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings

Related product manuals