EasyManuals Logo

Keithley DMM6500 User Manual

Keithley DMM6500
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
DMM6500
-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 2:
Performance verification
DMM6500-905-01 Rev. E June 2021 2-11
To verify ac voltage accuracy:
1. Connect the DMM6500 HI and LO INPUT connectors to the calibrator as shown in the following
figure.
Figure 2: Connections for ac voltage verification 100 mV to 100 V ranges
2. On the DMM6500, press the FUNCTION key and select AC voltage.
3. On the home screen, select the button next to Range and select 100 mV.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Make sure that detector bandwidth is set to 30 Hz.
AC voltage is specified for the detector bandwidth setting of 3 Hz. Three Hz measures accurately for
input signals from 3 Hz to 300 kHz, with reading rates ≈ 2 readings/s. To improve verification
throughput to ≈ 20 readings/s, set detector bandwidth to 30 Hz for frequencies of 30 Hz to 300 kHz.
To verify frequencies 1 kHz and higher, set the detector bandwidth to 300 Hz for faster ≈ 200
readings/s throughput.
7. Source ac voltages for each of the frequencies listed in the Verify the ac voltage 100 mV range
(on page 2-12) table.
8. Repeat these steps for each range and frequency listed in the tables below. For each voltage
setting, be sure that the reading is within low and high limits.

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings

Related product manuals