DMM6500
-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3:
Adjustment
DMM6500-905-01 Rev. E June 2021 3-19
Prepare your DMM6500 for a dc voltage adjustment
To prepare the DMM6500 for a dc voltage adjustment:
1. Connect a cable between the calibrator and the DMM6500 as shown in the figure below.
Figure 24: Connection for a dc voltage accuracy adjustment
2. Allow the DMM6500 to settle for five minutes.