EasyManuals Logo

Keithley DMM6500 User Manual

Keithley DMM6500
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
DMM6500
-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3:
Adjustment
DMM6500-905-01 Rev. E June 2021 3-23
Command parameters Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus value*
Reference
DMM dc
current range
cal_TS6
n/a
n/a
n/a
n/a
10 µA range (+) full scale nominal
cal_DCI_10uA_fs_pos
dc current
0.00022
0.00001
8508A or 8588A
Reading
10 µA
cal_DCI_DIGI_10uA_fs_pos
dc current 0.00022 0.00001 8508A or 8588A
Reading
10 µA
10 µA range (-) full scale nominal
cal_DCI_10uA_fs_neg
dc current 0.00022 -0.00001
8508A or 8588A
Reading
10 µA
cal_DCI_DIGI_10uA_fs_neg
dc current
0.00022
-0.00001
8508A or 8588A
Reading
10 µA
100 µA range (+) full scale nominal
cal_DCI_100uA_fs_pos
dc current
0.00022
0.0001
8508A or 8588A
Reading
100 µA
cal_DCI_DIGI_100uA_fs_po
s
dc current
0.00022
0.0001
8508A or 8588A
Reading
100 µA
100 µA range (-) full scale nominal
cal_DCI_100uA_fs_neg
dc current
0.00022
-0.0001
8508A or 8588A
Reading
100 µA
cal_DCI_DIGI_100uA_fs_ne
g
dc current
0.00022
-0.0001
8508A or 8588A
Reading
100 µA
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus value*
Reference
DMM dc
current range
1 mA range (+) full scale nominal
cal_DCI_1mA_fs_pos
dc current
0.0022
0.001
8508A or 8588A
Reading
1 mA
cal_DCI_DIGI_1mA_fs_pos
dc current
0.0022
0.001
8508A or 8588A
Reading
1 mA
1 mA range (-) full scale nominal
cal_DCI_1mA_fs_neg
dc current
0.0022
-0.001
8508A or 8588A
Reading
1 mA
cal_DCI_DIGI_1mA_fs_neg
dc current
0.0022
-0.001
8508A or 8588A
Reading
1 mA
10 mA range (+) full scale nominal
cal_DCI_10mA_fs_pos
dc current
0.022
0.01
8508A or 8588A
Reading
10 mA
cal_DCI_DIGI_10mA_fs_pos
dc current
0.022
0.01
8508A or 8588A
Reading
10 mA

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings