DAB-BA-e-1813 23
8.2 Calibrate / adjust temperature of the heating module
8.2.1 Temperature calibration
We recommend performing occasional tests of the instrument’s temperature value,
using the optional temperature calibration set KERN DAB-A01.
Before you do this, allow the device to cool down for at least 3 hours after the last
heating phase.
Preparation:
Switch off moisture analyser
Install temperature calibration set as per illustration
Turn on moisture analyser