15 Instant help
Possible causes of errors:
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then restart from the beginning.
Fault Possible cause
Display is not lit up.
• The appliance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
• Power supply interrupted.
• Fuse has blown
The display does not change
when a sample is being loaded
• Sample dish / dish holder is fitted incorrectly.
The weight display changes
does not appear.
• Sample dish has contact with windshield or
heating hood.
• Draught/air movement
• Table/floor vibrations
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Incorrect measuring result
• Check adjustment
• No resetting to zero before loading the sample
Measurement is taking too
long
• Incorrect shut-down criteria set
Measurement is not
reproducible
• Sample is not homogenous
• Drying time is too short
• Drying temperature too high (e.g. oxidation
sample material, boiling point of sample
exceeded)
• Temperature sensor soiled or defective
Drying does not start
• Heating hood open
• The mains supply connection has been
interrupted (mains cable not plugged in/faulty).