32 DAB-BA-e-1813
3. Selecting switch-off criterion
A switch-off criterion defines under which conditions the device should cancel the
drying process. Switch-off criteria save continuous time controls and manual
cancelling of drying processes. Furthermore they ensure that measurements are
always finished under the same conditions, thereby providing repeatable
measurements.
After acknowledgement of the selected heating profile by , the currently
set heating profile and switch-off criterion flashes.
Using arrow keys select required switch-off criterion and confirm
by .
Selectable settings suitable as switch-off criteria include:
<AUTO>
Automatic switch off
(2mg weight loss in 45 s)
This switch-off criterion is based on a
weight loss per time unit. As soon as the
medium weight loss drops below the
desired value per unit of time, the
measurement will be finished
automatically.
<TIMING>
Time-controlled switch off,
selectable 3 min 99 min
When this switch-off criterion is selected
the measurement will continue until the set
drying time has lapsed
<MANUAL>
Manual switch-off by pressing stop key