EasyManuals Logo
Home>KERN>Measuring Instruments>DAB 200-2

KERN DAB 200-2 User Manual

KERN DAB 200-2
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
DAB-BA-e-1813 33
4. Set drying temperature
After acknowledgement of the selected switch-off criterion by , the
currently set drying temperature is flashing.
Using arrow keys , select the desired value for the first or the first
two digits and confirm by .
Using arrow keys select the desired value for the last digit.
Confirm input by .
If as switch-off criterion <AUTO> or <MANUAL> is selected, the parameter
setting is finished here. All parameter settings will be applied to a subsequent
drying process and displayed on screen. All settings are stored under the ident
number <1> until a new method is allocated to the ident number.
<AUTO>
<MANUAL>
When <TIMING> is selected, the display of setting the drying period of time is
flashing, see step „5“.