EasyManua.ls Logo

KERN EW

KERN EW
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2. Table Increasing Calibration Weight For Linearity Adjustment
( ) : Cumulative total of weights on the scale
Model
Display
EW 220-3NM
EG 220-3NM
EW 420-3NM
EG 420-3NM
EW 620-3NM
EG 620-3NM
EW 820-3NM
on =
0g
(0g)
0g
(0g)
0g
(0g)
0g
(0g)
on 1
50g
(50g)
100g
(100g)
150g
(150g)
200g
(200g)
on 2
50g
(100g)
100g
(200g)
150g
(300g)
200g
(400g)
on 3
50g
(150g)
100g
(300g)
150g
(450g)
200g
(600g)
on 4
70g
(220g)
120g
(420g)
170g
(620g)
220g
(820g)
Calibration
Weight
Required
50g x 4
20g x1
100g x 4
20g x 1
100g x 4
50g x 4
20g x 1
200g x 4
20g x 1
Model
Display
EW 2200-2NM
EG 2200-2NM
EW 4200-2NM
EG 4200-2NM
EW 6200-2NM EW12000-1NM
on 0
0g
(0g)
0g
(0g)
0g
(0g)
0g
(0g)
on 1
500g
(500g)
1kg
(1000g)
1,5kg
(1500g)
3kg
(3000g)
on 2
500g
(1000g)
1kg
(2000g)
1,5kg
(3000g)
3kg
(6000g)
on 3
500g
(1500g)
1kg
(3000g)
1,5kg
(4500g)
3kg
(9000g)
on 4
700g
(2200g)
1,2kg
(4200g)
1,7kg
(6200g)
3kg
(12000g)
Calibration
Weight
Required
500g x 4
200g x 1
1kg x 4
200g x 1
1kg x 4
500g x 4
200g x 1
2kg x 4
1kg x 4
Please use the weight for F1 (OIML R – 111) or higher class to maintain the
accuracy.
- 21 - EW_EW-C_EG_N-SH-e-0722