6.4 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage.
Only use original KERN mains adapters. Using other makes requires consent by
KERN.
6.5 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
Provide adjustment weight.
The required adjustment weight depends on the capacity of the
weighing system. Carry out adjustment as near as possible to the
scale’s maximum weight. Info about test weights can be found on the
Internet at: http://www.kern-sohn.com
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.