KXS_KXG-TM-BA_IA-e-1932 47
SPEEd Reaction speed,
selectable 0, 1, 2, 3, 4, 5*, 6, 7, 8, 9
< 0 >: Sensitive/fast
< 9 >: Insensitive / slow
-nt
Negative tare: on-off*:
Function only to „on“, when the weighing system is not
used in the open sales outlets. Function permits reduction
of the tare weight, pressing the TARE-key.
Function only available, when P4 oth- St (Multitara) to „on“.
P2 CoM
Interface
parameter
ModE
St1*
One output for stable weighing value.
Return to zero needed between values.
St2
One output for stable weighing value. No
return to zero needed between values.
StC
Continuous data output of stable weighing
values
Pr1
An issue takes place after pressing the
PRINT key.
Pr2
Manual totalizing, see chap. 10.2
The weight value will be added into the summation
memory and issued after the M+-key was pressed.
Auto
For automatic totalizing see chap. 10.3
This function is used to issue and add individual
weighing values automatically to the summation
memory on unloading of weighing scale.
CoMAnd
For remote control commands, see chap.
13.4
Cont
Continuous data output
bAud
Available baud rate: 600, 1200, 2400, 4800, 9600*,
19200
Pr
E71
7 bits, even parity
o71
7 bits, odd parity
n81*
8 bits, no parity
rPS
Continuous data transfer
Select interval 1, 2, 4, 8 16 or MAX*
PtyPE
PtUP*
Standard printer setting
Lab
Lab x
For data output format see chap. 13.2
selectable 0*, 1, 2, 3
Prt
Prt x
LanG
ENG*
Standard settings English
rtC
Set RTC (date/time)
rtCFO
Set format for date, selectable
<year_month_day> or <day_month_year>
rS485
Enter ID for RS485 interface
bLAnK
off*
Print a blank line on output format
on