6.8.1 Adjustment with internal weight (only models PNJ)
The internal adjustment weight is available at all times for starting adjustment via
keyboard stroke.
Press the F-key and keep pressed until „Aut.CAL“ is
displayed.
With pressed TARE key, press the F-key, then release
both keys at the same time.
The motor noise of the loading system for the internal
adjustment weight can be heard, the internal
adjustment is started.
After successful adjustment the balance automatically
returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing pan) the display will show an error message,
repeat adjustment.
The adjustment protocol output is started upon
connection to an optional printer and activated GLP
function (see chap. 11.2.1).