6.8.2 Adjustment with external weight (only models PNS)
Carry out adjustment as near as possible to the balance’s maximum weight
(recommended adjustment weight see chap. 1).
Info about adjustment weights can be found on the Internet at:
http://www.kern-sohn.com
Press the F-key and keep pressed until „CAL“ is
displayed.
With pressed TARE key, press the F-key, then release
both keys at the same time.
Ensure that there are no objects on the weighing pan.
With displayon FS“ place the required adjustment
weight carefully in the center of the weighing pan.
The adjustment process is started.
example
After successful adjustment the balance automatically
returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing pan) the display will show an error message,
repeat adjustment.
Take away adjustment weight.