54 PNJ_PNS-BA-e-1713
10.2 Settings for Stability and Response
Exists the possibility to tune the stability of the display and the degree of reaction of
the balance to the requirements of certain applications or the environmental
conditions.
Please note that in general slowing down reaction times result in higher stability of
the set data handling, while speeding up reaction times have an influence on the
stability deterioration.
Installation site
Menu setting „5.rE.
Menu setting „6.S.d“
Quiet
Busy
1
1
2
2
3
3
4
4
5
Press the F-key and keep pressed until „Func“ is
displayed.
When releasing the button, the first function „1.b.G.„ is
displayed with the current setting.
Press F-key repeatedly until „5.rE.“ or „6S.d with the
current setting is displayed.
Use the TARE-key to select the desired setting.
Save / back to weighing mode:
Confirm setting with SET-key. The balance returns to
weighing mode.